SV7200
Vortex flow meter with display
SVR34XXXIRKG/US-100
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — SV7200-00 — 02.02.2023 —
1 G 3/4
DN 20
Product characteristics
Number of inputs and outputs
Number of digital outputs: 2
Measuring range 5...100 l/min 0.3...6 m³/h
Process connection
threaded connection G 3/4 DN20
Application
Special feature
Gold-plated contacts
Application
for industrial applications
Media
water; glycol solutions; coolants
Medium temperature [°C]
-10...90
SV7200
Vortex flow meter with display
SVR34XXXIRKG/US-100
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — SV7200-00 — 02.02.2023 —
Pressure rating [bar]
12
Pressure rating [Mpa]
1.2
Note on pressure rating up to 40 °C
MAWP (for applications
according to CRN)
[bar]
4.3
Electrical data
Operating voltage [V]
18...30 DC
Current consumption [mA]
< 30
Min. insulation resistance [MΩ]
100; (500 V DC)
Protection class
III
Reverse polarity protection
yes
Power-on delay time [s]
< 3
Inputs / outputs
Number of inputs and outputs
Number of digital outputs: 2
Outputs
Total number of outputs
2
Output signal
switching signal; frequency signal; IO-Link; (configurable)
Electrical design
PNP/NPN
Number of digital outputs
2
Output function
normally open / normally closed; (parameterisable)
Max. voltage drop switching
output DC
[V]
2.5
Permanent current rating of
switching output DC
[mA]
100
Short-circuit protection
yes
Overload protection
yes
Measuring/setting range
Measuring range 5...100 l/min 0.3...6 m³/h
Display range 0...120 l/min 0...7.2 m³/h
Resolution 0.5 l/min 0.02 m³/h
Set point SP 6...100 l/min 0.36...6 m³/h
Reset point rP 5...99 l/min 0.3...5.94 m³/h
Frequency end point, FEP 20...100 l/min 1.2...6 m³/h
In steps of 0.5 l/min 0.02 m³/h
Frequency at the end point
FRP
[Hz]
100...1000
Measuring dynamics
1:20
Temperature monitoring
Measuring range [°C]
-10...90
Display range [°C]
-30...110
Resolution [°C]
0.5
Set point SP [°C]
-9...90
Reset point rP [°C]
-10...89
In steps of [°C]
0.5
Frequency start point, FSP [°C]
-10...70
Frequency end point, FEP [°C]
10...90
SV7200
Vortex flow meter with display
SVR34XXXIRKG/US-100
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — SV7200-00 — 02.02.2023 —
Frequency at the end point
FRP
[Hz]
100...1000
Accuracy / deviations
Flow monitoring
Accuracy (in the measuring
range)
± 2 % MEW; (water)
Repeatability
± 0,5 % MEW
Temperature monitoring
Accuracy [K]
± 1
Response times
Flow monitoring
Response time [s]
1; (dAP = 0)
Damping process value dAP [s]
0...5
Temperature monitoring
Dynamic response T05 / T09 [s]
T09 = 6
Software / programming
Parameter setting options
hysteresis / window; normally open / normally closed; switching logic;
frequency output; switch-on/switch-off delay; Damping; Display unit
Interfaces
Communication interface
IO-Link
Transmission type
COM2 (38,4 kBaud)
IO-Link revision
1.1
SDCI standard
IEC 61131-9
Profiles
Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis
SIO mode
yes
Required master port type
A
Process data analogue
2
Process data binary
2
Min. process cycle time [ms]
3
Supported DeviceIDs Type of operation DeviceID
Default 492
Operating conditions
Ambient temperature [°C]
0...60
Note on ambient temperature medium temperature < 80 °C
medium temperature < 90 °C: 0...50 °C
Storage temperature [°C]
-20...80
Protection
IP 65; IP 67
Tests / approvals
EMC DIN EN 61000-6-2
DIN EN 61000-6-3