Liquid Detection Fiber Sensor easy enough to
use for anyone in universities, research
laboratories and production fi eld
Output 1
(Dark-ON)
Detection of liquid in tube ON ON OFF
Output 2
(Light-ON)
Detection of tube OFF ON ON
*Please refer to the instruction manual for setting method of fi ber amplifi er.
No specialized technician required for the installation and removal of tube
When the Liquid Detection Fiber is used in combination with the 2-output fi ber amplifi er (FX-502, FX-502P),
the tube / liquid condition can be indicated in three patterns.
Liquid Detection Fiber accepts tubes that
are commonly used with cell culture
apparatus.
Mounting holes are designed to allow the sensor installation
in two different orientations so that the tube removal
direction can be selected.
● Dedicated tool and cable ties are required for the
replacement of tube.
● Risk of tube damage if the tube is replaced by a
person not familiar with the replacement procedure.
● The tube and sensor must be secured in place by
tightening the cable ties with appropriate tightening
force to prevent tube deformation.
● Sensitivity adjustment must always be made after
tube replacement.
New systemConventional system
Simple one-touch
system for anyone
Need Cable Tie
Two wide bracing arms
hold fl exible tube securely.
Install the sensor vertically
for upward removal of tube.
Install the sensor horizontally
for sideway removal of tube.
2-output fi ber
amplifi er
• FX-502
• FX-502P
No tube With tube
and liquid
With tube
but no
liquid
Tube removalTube installation
Diffi cult to replace tube
* If the sensing is unstable due to the deviations of material or diameter of the
replacement tube, adjust the amplifi er’s threshold setting and confi rm stable
sensing before using.
One-touch system for installation and removal of tube
Monitoring for prevention of nonattachment of tube
Applicable tubes:
Silicone and PVC tubes
Two selectable removal direction
Upward removal of tube
Sideway removal of tube
Applicable tubes
Material Diameter (Outside / Inside)
Silicone / PVC
ø4 × 2 mm
ø5 × 3 mm
ø6.4 × 3.2 mm (
1/4 × 1/8 in)
ø9.5 × 6.4 mm (
3/8 × 1/4 in)
Bracing
arms
Cable tie
Anti-skid
tube
Hold the tube and
lift it straight up.
Push in the tube
straight from above the
bracing arms.
Bracing
arms
2