SM9601
Magnetic-inductive flow meter
SMN32XGXFRKG/US-100
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — SM9601-00 — 27.07.2023 —
Medium temperature [°F]
14...176
Pressure rating [bar]
16
MAWP (for applications
according to CRN)
[bar]
16
Electrical data
Operating voltage [V]
18...32 DC; (to SELV/PELV)
Current consumption [mA]
< 150
Protection class
III
Reverse polarity protection
yes
Power-on delay time [s]
5
Inputs / outputs
Number of inputs and outputs
Number of digital outputs: 2; Number of analogue outputs: 1
Inputs
Inputs
counter reset
Outputs
Total number of outputs
2
Output signal
switching signal; analogue signal; pulse signal; frequency signal; IO-Link; (configurable)
Electrical design
PNP/NPN
Number of digital outputs
2
Output function
normally open / normally closed; (parameterisable)
Max. voltage drop switching
output DC
[V]
2
Permanent current rating of
switching output DC
[mA]
250; (per output)
Number of analogue outputs
1
Analogue current output [mA]
4...20; (scalable)
Max. load [Ω]
500
Analogue voltage output [V]
0...10; (scalable)
Min. load resistance [Ω]
2000
Pulse output
flow rate meter
Short-circuit protection
yes
Type of short-circuit
protection
pulsed
Overload protection
yes
Frequency of the output [Hz]
0.1...10000
Measuring/setting range
Measuring range 80...4800 gph 1.3...80 gpm
Display range -5760...5760 gph -96...96 gpm
Resolution 5 gph 0.1 gpm
Set point SP 105...4800 gph 1.7...80 gpm
Reset point rP 80...4775 gph 1.3...79.6 gpm
Analogue start point ASP 0...3840 gph 0...64 gpm
Analogue end point AEP 960...4800 gph 16...80 gpm
Low flow cut-off LFC < 240 gph < 4 gpm
In steps of 5 gph 0.1 gpm
Measuring dynamics
1:60