Quality, Safety, Responsibility
Datasheet
Carbon Monoxide
sensor
SGX Solid Polymer Electrolyte Gas Sensors
The SGX series of PS1 and PS4 Electrochemical gas sensors are using a revolutionary ‘Solid Polymer
Electrolyte’ technology that is based on the principle of catalytic reaction. The target gas to be measured
generates a very small current, proportional to the gas concentration. Our technology oers a stable, high
quality and cost-eective manufacturing process. The SGX solid polymer electrolyte gas sensors are
available in a very small size, are highly sensitive, do not use power and have very low cross sensitivity
from other gases.
PS1-CO-1000
PS4-CO-1000
PS1-CO-1000
PS4-CO-1000
Technical Specications
25 ± 10 nA / ppm
Sensitivity
Measurement Range
0 – 1000 ppm
Zero Current ± 100 nA
Maximum Overload 2000 ppm
Response Time
T50 < 10s,
T90 < 30s
Repeatability < 1%
Lower Detectable Limit (LDL) ≤ 2 ppm
Linear Range 1000 ppm
Resolution (16Bit ADC) 0.1ppm
Performance
Operating Humidity Range 15-95% RH
Pressure Range 800 to 1200 hPA
Storage Temperature
0 to 20
Temperature Range
-40°C to +55°C
Environmental Details
Operating Principle
Amperometric,
3-electrode
Recommended Load Resistor 100 Ω
Bias Voltage 0 mV
Warm Up Time
< 60 s
Operation
Long-Term Drift
< 1 %/month
Zero Drift in Clean Air
< 2 ppm
Storage Life
12 months
Expected Lifetime
> 5 years in air
Storage conditions
0-20°C
Warranty
12 months
Lifetime Details
Housing Material PPO
Housing
Important Notes
All performance is based on conditions at 20°C,
50% RH and 1 atm, ow rate>150qcm/min,
using SGX recommended circuitry.
Sensor performance is temperature dependant;
please contact SGX for temperature
performance other than 20°C.
Do not solder to the connector pins as this may
damage the sensor and thereby invalidate the
warranty.
Details on recommended connector pins can be
found in the Frequently Asked Questions within
the Gas Sensor section of the SGX website.
Features
Small size
Wide teperature range
Low cross sensitivity
No electrolyte leakage
High sensitivity
Fast response time
Extreme linear response up to high
concentration
SGX Europe Sp. z o.o.
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
T: +48 (0) 32 438 4778
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
© SGX Sensortech 2012-2022 DS-0368-PS1&PS4-CO-1000, Issue 1, 23-Jan-2022, Page 1/3
PS1-CO-1000 < 0.7g
PS4-CO-1000 < 6g
Weight
Key applications
TLV Monitoring
Oil & Petrochemical Industry
Leak detection
Steel Industry
Parking Garages
SGX Europe Sp. z o.o.
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
T: +48 (0) 32 438 4778
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
© SGX Sensortech 2012-2022 DS-0368-PS1&PS4-CO-1000, Issue 1, 23-Jan-2022, Page 2/3
Dimensions
PS1-CO-1000
PS4-CO-1000
Cross Sensitivity
Gas
Formula
Test Concentration
Sensor Reading
Alcohols R-OH
NH
3
CO
2
CL
2
C
2
H
4
HCN
H
2
H
2
S
C
3
H
7
OH
CH
4
NO
NO
2
SO
2
1000ppm 0ppm
Ammonia
50ppm 0ppm
Carbon Dioxide
Chlorine
1000ppm
1ppm
0ppm
0ppm
Ethylene
Hydrogen Cyanide
10ppm
50ppm
0ppm
0ppm
Hydrogen
100ppm 20ppm
Hydrogen Sulphide
50ppm 0ppm
Isopropanol
Methanal
Nitric Oxide
Nitrogen Dioxide
Sulphur Dioxide
1000ppm
1%vol
25ppm
10ppm
10ppm
0ppm
0ppm
n.e
0ppm
0ppm
Note:
1) The above interference factors may vary due to different sensors and service life, please refer to the actual test results.
2) This table is not complete for all cross gases, other gas please contact with us.
C
W
R